VM-100C Portable Ventilator

  • Detail
  • Parameters

Technical Specifications

Physical Specifications

Dimension: Width 300mm, Depth 160mm, Height 230mm,Weight 3Kg

Carrying Packages Dimension: Width 460mm, Depth 200mm,Height 400mm, Weight 7Kg

Packaging Dimension: Width 620mm, Depth 240mm, Height410mm, Weight 11Kg

Storage/Shipping Requirements

Temperature: -20 to 70°C

Relative Humidity less than 93%

The atmosphere pressure: 70 to 106 kPa

Operating Requirements

Temperature: 5 to 40°C

Relative Humidity less than 80%

The atmosphere pressure: 70 to 106 kPa

Electrical Specifications

Input Voltage Requirements: AC Voltage 110V~240V; DCVoltage 12V

Inlet Battery: 12V, 2Ah (Can be used for 8 hours)

Frequency: 50Hz/60Hz

Power: 2A

Pneumatic Specifications

Pressurized Oxygen: 0.4MPa medical oxygen inlet oruse 2L oxygen cylinder (Choice of suppliers)

Performance Specifications

Power: ON/OFF

Modes: CMV(IPPV), A/C(SIPPV), SIMV, SPONT, PEEP(Choice of suppliers)

Tidal Volume: 50~1200ml

Breath Rate: 5~60 bpm

PEEP: 0~2kPa

O2%: 45~90%

Sigh: 1~10 per 100 breath

Peep: 10-60hpa (Choice of suppliers)

Trigger Level: -10 ~ 0hpa

Over Pressure Relief: <6kpa

Alarms and Indicators

High Pressure Alarm Limit: 2~6kpa flash, redindicator

Low Pressure Alarm Limit: 0. 2~2kpa Red indicator

Alarm Silence: 120sec

Standard Accessories

Ventilator                       1unit

Oxygen Cylinder                1pcs (2.5L can be used for 1hour)

Breath Circuit                   1pcs (1.2m)

Carrying Package               1pcs (all parts in it)

Breath Value                    1pcs

Mask                           1pcs

Charger                        1pcs

Battery                         1pcs (can be used for8 hours)

Wrench                        1pcs

Oxygen Pressure Reducer       1pcs (for Normal Oxygen Cylinder)

Lung Simulators                 1pcs

Hand Book                      1pcs

CD For Operation                1pcs


Home
Customer service
Add to Enquiry List