Introduction
This isthe first manual, light-weight, compact, portable Anesthesia Machinespecifically
designedfor surgical locations where space or access is limited, such as:
• OfficeBased Anesthesia (OBA)
• Dentalanesthesia
•Missionary off-shore anesthesia
• Remotelocations in hospitals outside O.R.
•Disaster scenes and mass casualty incidents
•Front-line military hospitals
•Shipboard, aircraft
Anylocation where size, mobility and portability are important.
Technical Parameter
1. Workingmode: Manual
2.Patient scope: from Pediatric to Adult (3kgs over)
3. Tidalvolume: 20-1500ml
4.Working pressure: O2:0.3—0.4Mpa, N2O:0.3—0.4Mpa
5.Flowmeters: 2 tube 2 gas flowmeter; O2 range: 0.1-10L/min, N2O range:0.1-10L/min.
6. O2 andN2O linkage device (ORC system), FiO2 ≥25%
7. APLvalve adjustment range: 0.5-6.0Kpa
8. O2Flush: 30-60L/min
Standard Configuration
- MainUnit
- OneVaporizer (choose one from below vaporizers)
Smallvaporizer: Halothane, Isoflurane, Enflurane
Bigvaporizer: Halothane, Isoflurane, Enflurane and Sevoflurane.
-Standard Accessories of breathing tubes for connecting and Mask etc.
-In-built Co2 absorber Canister with capacity of 2kg.
Two Options
①Anesthesia Ventilator TH-1A
Special Feature
- IndependentAnesthesia Ventilator, with IPPV and Manual mode
- Withnecessary monitoring and alarm function,
to makesure the operation safety
- Gassupply: O2, or Air
- Withinternational connector, can be matched with
variousbrands of anesthesia machines
Specification
-Respiration mode: IPPV, SIMV, Manual
-Respiratory frequency: 1 to 65BPM
- Display:5.5´ SNT screen
- I:Eratio: 8:1~1:8
-Waveform: P-t, V-t
- MonitorParameters: Tidal volume; Minute Ventilation; Peak
- Tidalvolume: 50ml~1600ml
② Anesthesia VentilatorTH-1
Special Feature
- IndependentAnesthesia Ventilator, with IPPV and Manual mode
- Withnecessary monitoring and alarm function,
to makesure the operation safety
- Gassupply: O2, or Air
- Withinternational connector, can be matched with various
brands ofanesthesia machines
Specification
- Controlmethod: Gas driving electronic controlled
- Breathfrequency: 1~90BPM
- Display:High light LED display
- I:Eratio: 4:1~1:6
-Respiration mode: IPPV, Manual
- MonitorParameters: Tidal volume; Minute Ventilation; Breath Frequency; Airway Pressure
- Tidalvolume: 50~1500ml