AM-3 Anesthesia Machine

  • Detail
  • Parameters

Technical Specifications

PARAMETER

SPECIFICATION

Main Body

High-strength engineering   plastic rack, light, beautiful and corrosion resistant

Scope of Application

Adult, child

Gas Source

O2: 0.27--0.55Mpa;

N2O: 0.27--0.55MPa.

Flow Meter

O2:   0.1--1.0L/min; 1.1--10L/min

N20:   0.1--1.0L/min; 1.1--10L/min

O2, N2O Linkage and N2O

Stopper

When using nitrous   oxide, oxygen concentration>25%;

When the oxygen   pressure < 0.2 MPa, the flow of nitrous oxide would be cut off.

Flow Rate of Rapid   Oxygen Supply

25--75L/min

Low Oxygen Pressure   Alarm

There will be sound   alarm when the oxygen pressure < 0.2MPa

Vaporizer

It has the function of   automatic compensation based on pressure, temperature, and flow rate. The   regulation range of evaporator concentration is 0-5vol%.

Among Halothane,   Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as   required by the customer. And vaporizers that are imported with original   packaging are also available.

Respiratory Circuit

Working mode:   all-close, semi-close, semi-open;

Releasing pressure:   0.5-7kPa.

Respiratory Bellows

Bellows for adults,   bellows for children, tidal volume range:0-1500 ml.

VENTILATOR

Display Mode

10.4” Touch Screen,   High-definition TFT color LCD.

Ventilation   Mode

IPPV, SIPPV, SIMV,   PEEP, MANUAL, SIGH (IMV, VCV, PCV)

Ventilation Parameters

Tidal   volume           50--1500ml

Rate                  1--99bpm

SIMV rate             2--20bpm

I:E                    2:1--1:8 (4:1--1:8)

Inspiratory trigger   pressure   -1.0--2.0kPa

PEEP                 0--2.0kPa

Pressure range       1.0--6.0kPa

SIGH         1.5 times the inspiratory time   (70-150 adjustable)

Inspiratory   platform       0--50%

Parameters for   Ventilation

Monitoring

Tidal volume,   ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak   pressure of airway, pressure - time waveform, flow rate – time waveform,   PEEP, inspiratory trigger pressure, Inspiratory platform

(autonomous inspiratory   rate, lung compliance, airway resistance, flow-volume loop, pressure-volume   loop)

Monitoring of Oxygen

Concentration

21%--100%

SECURITY   ALARM SYSTEM

Oxygen Concentration alarm

Upper limit setting   range   21%--100%;

Low limit setting   range     10%--80%.

Airway Pressure Alarm

Upper limit setting   range   0.3--6.0kPa;

Low limit setting   range     0.2--5.0kPa.

Per-minute Ventilation   Volume Alarm

Upper limit setting   range   3.0--30L/min;

Low limit setting   range     1.0--10L/min.

Sustained High-pressure

Alarm

It will give alarm when   stress have consistently been higher than 2.5 kPa

Suffocation Alarm

It will give sound and   light alarm if there is no tidal volume input for 15 seconds

POWER   ALARM

Power

AC 220V±10%, 50Hz±1Hz, (Storage battery,   UPS)

Mechanical Arms (End-expiratory   CO2 monitoring)

Note: Items within   brackets are optional.


Home
Customer service
Add to Enquiry List